02 November 2026

Automating high-throughput NMR

NMR was the first practical method used to identify fragments, described thirty years ago in this famous SAR by NMR paper. That headstart made NMR the go-to method for finding fragments, and as recently as ten years ago it was still in the lead, according to our poll in 2016. But as other methods have become faster and more effective, the relative popularity of NMR has declined, to fourth place in our 2024 poll. Now, a new paper in J. Med. Chem. by Bradley Falk and colleagues at Bristol Myers Squibb reports improvements that might help it move up the ranks.
 
One of the challenges of NMR is that it requires so many steps. Protein-detected NMR requires isotopically labeled protein, but even ligand-detected NMR involves measuring spectra for small molecules in the absence and presence of protein. The specific NMR settings may need to be tweaked for any given molecule, and the spectra are then analyzed, usually manually, to decide whether the molecules are likely binders.
 
The new paper focuses on automating virtually everything in ligand-detected NMR, from sample preparation through NMR runs to analysis. It is meant to be more general than a fragment screen with a curated library where each member is already well understood. First, novel compounds are examined in a band-Selective Optimized Flip Angle Short Transient (SOFAST) NMR experiment, which is particularly sensitive (down to 20 µM ligand). Testing in both DMSO-d6 and assay buffer provides solubility and quality control data. This is an essential step: as we noted some time ago more than a third of compounds in some vendors’ catalogs fail these quality checks. In addition, a CPMPG SOFAST experiment in assay buffer reveals whether the compounds aggregate, a hazard we wrote about here.
 
Next, an automated workflow funnels appropriate compounds to the next step, in which compounds are assessed for protein binding using CPMG or STD NMR. Four separate samples are tested: low or high concentrations of compound in the absence or presence of protein. Again, automated approaches are used to choose the specific NMR parameters and analyze the results. As the researchers note, “human input is reserved for evaluations that require integrating multiple lines of evidence.” The output places compounds into three bins, “advance to med chem,” “advance with orthogonal confirmation,” or the garbage bin.
 
The researchers describe an example using the model protein human carbonic anhydrase II (hCAII), which we’ve written about previously. The known ligand benzenesulfonamide was tested and passed into the “advance to med chem” bin, despite the fact that, with a KD of 0.8 µM, it is more potent than most ligands studied by NMR.
 
One thing missing from the paper is a summary of overall time and throughput. The experimental section states that the CPMG-SOFAST experiment itself takes half an hour, and indeed the researchers note that the platform is “instrument-time limited rather than sample preparation- or data analysis-limited.” In terms of absolute throughput, it probably can’t compete with other biophysical methods such as SPR. The automation and automated decision-making is the real advance. In 2019 I satirically imagined a fully autonomous drug discovery engine. As artificial intelligence becomes ever more intelligent, that post may prove prescient. If so, it will feature automated tools such as those described in this paper.

01 April 2019

Machines, fixing human disease

Last year we highlighted the secretive juggernaut DREADCO's move into drug discovery. Today they announced the launch of their new division SkyFragNet (not to be confused with the European graduate training program FragNet). Its audacious mission: “to eradicate human disease."

SkyFragNet will automate every aspect of drug discovery. The approach starts with a powerful docking method, in which all 166 billion members of GDB-17 will be docked against a target of interest. Synthetic schemes for the virtual hits will be computationally generated, and the compounds will be synthesized using automated flow synthesis and mass-directed purification.

Fragment hits that confirm in a panel of biophysical techniques will then undergo computational-based growing; SkyFragNet incorporates the latest AI algorithms to maximize the likelihood of success. As with the fragments, designed molecules will be synthesized and tested, first in biochemical and then in cell-based assays.

Although the folks at Mordor State College are trying to make animal testing obsolete, SkyFragNet will still rely on pharmaokinetic and pharmacodynamic studies. However, they have built a fully mechanized vivarium run entirely by robots - think of The Matrix but with mice in place of humans.

Finally, compounds that make it through this gauntlet will be scaled up under GMP conditions (automated, of course) for clinical trials. It remains to be seen how many compounds SkyFragNet will take into the clinic, or whether the success rates will be higher than those of their human counterparts.

Of course, with all this power comes enormous responsibility. If things go wrong, hopefully DREADCO will have the wisdom to Terminate the program. Eradicating human disease could be done in two very different ways.

16 July 2018

Rise of the machines for fragment optimization

Our latest poll (please vote on the right-hand side of the page!) is about fragment libraries. Once you have your library, you can screen it using a variety of approaches. But what do you do once you get hits? Computational methods are increasingly being adopted; just this year we’ve discussed two approaches: growing via merging and AutoCouple. A new paper in J. Med. Chem. by Philippe Roche, Xavier Morelli, and collaborators at Aix-Marseille University and several other institutions describes a method that combines virtual screening with automated real-world synthesis in a platform called diversity-oriented target-focused synthesis (DOTS).

The process is best described with an example, and the test case presented is the first bromodomain of BRD4, BRD4(BD1). The researchers, who had previously identified a xanthine-containing series of inhibitors, pared this back to fragment-sized compound F1. Crystallography revealed a nearby pocket, which the researchers attempted to target with DOTS.

The researchers built a virtual library of 576 sulfonamides extending off the para position of the phenyl ring of compound F1. These were then virtually screened against BRD4(BD1) using the S4MPLE molecular modeling tool in which the F1 portion was constrained in the crystallographically observed conformation while the variable bits were allowed to move. The 100 top-scoring molecules were examined more closely, and 17 representatives were chosen to be synthesized on an automated robotic platform. This was actually a fairly modest set, as the Chemspeed system they used can run up to 96 parallel reactions. The crude products were then tested in a fluorescence assay, and all of them showed improved activities compared to the initial fragment. The majority, such as compound 17, showed high nanomolar inhibition.

The 13 submicromolar compounds were then resynthesized, purified, and validated in thermal shift and isothermal titration calorimetry (ITC) assays; these orthogonal methods confirmed their activities. The crystal structure of compound 17 bound to BRD4(BD1) was also solved, and this revealed that – as designed – the initial fragment retained its binding mode while the added portion makes new interactions with the protein.






The fact that 14 of the 17 molecules synthesized were at least an order of magnitude more potent than the initial fragment is satisfying, though it is worth noting that bromodomains are not the most difficult targets. Also, all of the new molecules have lower ligand efficiencies than the initial fragment. Still, advances and combinations of computational and robotic approaches will certainly increase the throughput of synthesis and testing, and I expect to see more of these examples.

21 November 2016

New tools for NMR

As most of you know, Teddy has retired from active blogging, which is unfortunate not just for the loss of his wit but also for the loss of his expertise – particularly regarding NMR. But you blog with the army you have, not the army you want, so I'll take a stab at two recent papers on the subject.

The first, published in J. Med. Chem. by Chen Peng and colleagues at software maker Mestrelab in collaboration with Andreas Lingel and colleagues at Novartis, describes an automated processing program for just about any type of ligand-observed NMR data. After going into some detail on how “Mnova Screen” works, the program was benchmarked on three experimental data sets (on undisclosed proteins) which had previously been processed manually. The first was 19F data from a collection of 19 mixtures of up to 30 fluorinated compounds each – 551 altogether. Here the program performed quite well, identifying 56 of the 64 hits identified manually and misidentifying only 4 compounds as hits. Most of the false positives and false negatives were close to the predetermined cutoff threshold, which can be set as stringent or lax as desired.

T1ρ and STD NMR experiments on 55 individual protein-compound complexes were also examined, and the results were similarly positive. Of course, single compound experiments are easy to analyze, and the real test was with a set of 1240 compounds in 174 pools. Here the program was not quite as good, missing 16 of the 50 manually identified hits and coming up with 74 hits that had not been identified manually. Although most of these were false positives, closer inspection revealed that 10 of them are probably real. Moreover, some of the “false negatives” should perhaps not have been classified as hits in the first place. Clearly the program isn’t perfect, but it does seem to be a quick way to triage large amounts of data.

Of course, ligand-detected NMR provides at best only limited information on binding modes, which is where the second paper comes in, published in J. Biomol. NMR. by Mehdi Mobli (University of Queensland), Martin Scanlon (Monash University) and collaborators at Bruker and La Trobe University. The researchers were interested in finding inhibitors of the bacterial protein DsbA, and a previous screen had identified a weak fragment that initially proved recalcitrant to crystallography.

One of the best methods to determine the binding mode of a ligand is to look at intermolecular NOEs, NMR signals which only show up when two atoms are in close proximity to one another. In theory you can look at NOEs from ligands to the backbone amide protons in proteins, but this is technically challenging for aromatic ligands, of which there are many. Proteins have plenty of methyl groups – so many in fact that it can be difficult to correctly assign each methyl group to a specific residue, leading some researchers to only focus on isoleucine, leucine, and valine (ILV). However, by carefully studying more than 5000 high-quality protein ligand complexes, the researchers found that looking at all the methyl groups in a protein (ie, including those found in alanine, threonine, and methionine) greatly increases the number of protein-ligand complexes suitable for analysis.

The researchers were able to assign most of the methyl groups in DsbA using several approaches, and this allowed them to identify 11 NOEs between their ligand and ILV methyl groups. Modeling was unable to provide a unique binding mode, but by including 8 more NOEs to threonine and methionine methyl groups a single binding mode for the ligand was determined. Crystallography came through in the end too and confirmed the NMR-derived model.

Teddy would normally end his NMR posts by stating – often forcefully – whether he thought the tools under discussion were practical or not. NMR is one of the most popular methods out there, so new tools are clearly welcome. Since I'm no expert on the subject, I'll ask readers to weigh in – what do you think?

10 October 2016

Tips for high-throughput crystallography

X-ray crystallography is tied for second place among methods used in fragment-based lead discovery, according to our most recent poll. This makes sense, since structures are usually essential for advancing fragments to leads. Faster fragment-finding methods are usually used to triage fragments down to a manageable number of hits to feed into crystallography, but the high incidence of false negatives means that promising fragments might be inadvertently discarded. If structures are key goals at the end of a fragment screening campaign, why not start directly with crystallography?

In fact, this is exactly what more and more groups seem to be doing. The problem, historically, has been throughput. Increasing automation has been solving some of the mechanical issues (such as mounting crystals and collecting data at a synchrotron), but what about the actual processing? A recent paper in Structure by Andreas Heine and collaborators at Philipps-University Marburg and Helmholtz-Zentrum Berlin für Materialien und Energie provides some useful advice.

The protein in question is endothiapepsin, a model aspartic protease that is easy to crystallize and diffracts to high resolution. Earlier this year, we discussed the researchers’ work soaking 360+ fragments against this protein, and a companion paper gives detailed information on how several dozen fragment hits bind. The Structure paper describes an automated refinement pipeline, and highlights some of its most important features.

Determining a crystal structure involves iterative cycles of modeling the protein backbone and side chains into regions of “electron density.” One risk is “model bias,” illustrated memorably in this brief video. This is especially important for small molecules: since they represent such a tiny fraction of the overall structure, it is especially easy to see what you want to see. To avoid this, people often look for regions of electron density – which in addition to a bound small molecule could represent co-solvents, buffer, or an amino acid side chain that has unexpectedly moved – before doing much refinement.

The problem is that the electron density might be very spotty and easy to overlook. This is especially true for fragments that bind weakly and which are small by definition. Some initial refinement can thus improve the quality of the electron density maps. The researchers find that adding water molecules and including these in the refinement is the single most important step. Adding bound hydrogen atoms to the protein model is also helpful: even though each hydrogen only contributes one electron to the overall density, there are more than enough to make a meaningful difference. Finally, for very high resolution structures (better than 1.5 Å), it can help to treat each atom of the protein individually (anisotropic refinement of B factors, or atomic displacement parameters). However, at lower resolution, doing this can lead to overfitting. Incorporating these steps into the automated process revealed that 25% of fragments would have been missed had conventional methods been used.

The paper includes lots more detail that will be of interest primarily to crystallographers. Moreover, the data for all 364 fragment soaks has been uploaded to the protein data bank. This is a very high-quality data set: all the crystals diffracted to better than 2.0 Å resolution, with the mean being 1.35 Å, and should be a useful resource for those of you establishing your own automated processing system.

04 March 2015

Way Down in the X-Ray Weeds

So, what I know about the details of crystallography can fit on the head of a pin...a small pin.  You put pure protein in multiwell plates and then do a huge matrix of crystallization conditions until tiny little crystals form.  Big crystals are best, but you can use tiny crystals or seeds, or with recent advances in technology, to actually collect data.  Then, through some wizardry (some sort of inverse transform) you make spots go to electron density, then with will power and what used to be SGI machines, you thread your protein sequence in, et voila a model of the structure.  I typically don't go for methods papers in fields I have almost no clue in, but this one intrigued me.

This paper aims to increase the efficiency of soaking fragments into crystals to take advantage of 3rd generation synchrotrons. These machines/labs/setups/doohickeys use acoustic droplet injection (ADE), which many people may already be aware of.  In this approach, each fragment soaks into a protein crystal either directly on data collection media or on a moving conveyor belt which then delivers the crystals to the X-ray beam.  The source of inefficiency comes from the time required to soak the fragment in to the crystals (for those where the apparatus is inside the X-ray station. I have no idea what that means, but here is a google image search that might give you an idea.)  A second source is the limit of evaporative dehydration during the fragment soak.  

Using the model system lysozyme and thermolysin the identified factors which can increase efficiency: namely smaller crystals can be used to decrease the soak time.  By small crystals, they are talking things that are 100 microns or less.  The authors go on to state that:
These techniques efficiently use fragment chemicals (~2.5 nL per screened condition), protein (~25 nL per screened condition), space (1120 screened conditions per standard shipping Dewar; no limits using a conveyor belt), and synchrotron beam time (less than 1 second/screened condition).  Evaporative dehydration of the protein crystal limits these fragment screening applications to systems where the fragment soak time is not prohibitive. Slow-binding compounds can be screened (without time constraint) in trays using ADE, but will consume significantly more resources such as purified protein and chemical compounds (~1 µl per screened condition). Hence, it is desirable to identify promising cases where the cost-efficient on-micromesh or on-conveyor soaking methods are adequate.
So, what did I really get out of this paper.  I am amazed by the miniaturization and automation that exists in synchrotrons.  It is really amazing. Its good to get and read the literature for another field.  It can be enlightening.  If I am looking at this correctly, they can screen a 1000 fragments with ~1mg of protein.  With that said, how many people screen for fragments in this way?  It seems not to be resource intensive, if you are sitting at a synchrotron.  But, how much does it cost to sit at that synchrotron?  Are the problems called out here something people encounter every day, or is this a "First World problem" for those sitting at synchrotrons.

What really got me was that the majority of the authors are high school students and undergraduates. This emphasizes to me the commoditization of X-ray, and really all services.  There is a very high level of training that goes in to solving the structure; I get that.  But it seems that many of the steps are commmodities, if you will.  When I was at Merck, they had a directive called (in some form) I-C-E: Innovation-Commoditization-Experimentation.  The concept was the highly trained (and highly paid) scientists needed to focus on innovation.  Once innovation was achieved it led to experimentation (figuring out how to run it routinely).  After that it was a commodity and should be outsourced to enable those scientists to go back to innovating.  It makes sense from a business standpoint, but scary from the scientists standpoint.  I am all for full employment for scientists in industry (trust me on this), but outsourcing can co-exist in industry. Look at the growth in providers from 2011 to 2014. Not sure where I am going with this, but food for thought.

17 March 2014

This is another way to do it.

The key to doing something right is to following the directions.  How closely you follow the directions, or don't follow, can be the difference between brilliance and just a good performance, e.g. cooking.  Sometimes, directions are meant as guidelines, like the Pirate Code or the Voldemort Rule.  Late last year, and blogged about here, I published a paper in Current Protocols on how to prosecute an STD screen.  A recent paper in PLOSOne, shows how someone else runs their screens, but with details on library construction, solubility testing, and more.  What makes this paper of interest is the level of detail that they provide.

Library Design: They assembled a diverse fragment library with the following rules: 110≤ molecular weight ≤350, clogP≤3, number of rotable bonds ≤3, number of hydrogen bond doners ≤3, number of hydrogen bond acceptors ≤3, total polar surface area ≤110, and logSw (aqueous solubility) ≥ −4.5. 
I am little confused by the figure and what the text says.  In the text, they seem to have relaxed the MW cutoff, but the figure shows that anything not Voldemort Rule compliant is tossed.  They also preferred that the compound has at least one aromatic peak (for easier NMR detection).  They purchased 1008 from Chembridge, solubilized at 200 mM in DMSO-d6 (ease of NMR detection, again) and then tested the solubility at 1 mM in water.  I would have added some salt here, 50 mM, but that is a quibble.  For purity, they claim a low level of impurity (< 15%)!!!  To me, this is a whole lot of impurity.  But, as has been noted here, purity levels vary from library to library.
Solubility Testing:  They then made sure to experimentally test every fragment for solubility.  I can agree more emphatically with this approach.  Bravo!  They go into great detail, which I will not attempt to replicate here, but thanks to open access, they have included the scripts in the supplemental.  Acceptable compounds had > 0.1 mM aqueous solubility.  For me, this is too low, but to each their own.  They ended up with 893 total fragments (89% passed).  The real data I would like to see is how many fail if the cutoff is set at 0.5 mM or higher.  
Pooling: They then describe their pooling strategy.  I like open access articles for a lot of reasons, and tend to overlook small editorial problems (typos, grammar, etc.), but in this case, let me rant.  The authors state in the text that a random mixing of compounds would lead to severe overlap, exemplified in 3a.  To me, it does no such thing. 

Their approach is very similar to the Monte Carlo-based one that has previously been discussed on this blog.  Their final pools contain 10 fragments at 20 mM (I assume in 100 % DMSO-d6). 
Screening: They also acquired the 1H spectrum, STD (-0.7 ppm, > 1 ppm from any methyl), and WaterLOGSY spectrum of every pool for future reference.  This is a very clever approach as the STD should give no signal while the WaterLOGSY should give inverted peaks for all compounds in the pool (when interacting with a target they will be "right-side up").  Again, the figure may show that (I think if you blow up the figure the WaterLOGSY spectra does have peaks) but it is very difficult to see. 
Three of the 90 pools (3.3%) showed peaks in the aromatic region, most likely due to aggregation (they observed precipitation).  I would like to know if those compounds showed STD peaks also had those methyl groups within 1 ppm of the saturation frequency.  I would also like to know if they removed those compounds from the library, or just dealt with it.  For a paper with a great level of detail, it falls flat in this respect.  
Screening is performed at 10uM Target: 500uM ligand and the following parameters: acquisition time of 1 s, 32 dummy scans, and relaxation delay of 0.1 s, followed by a 2 s Gauss pulse train with the irradiation frequency at −0.7 ppm or −50 ppm alternatively. The total acquisition time was 15 minutes with 256 scans.
Screen Analysis: One of the first things they noticed was that there were difference between the reference spectra (plain water) and the screening sample (protein buffer).  They decided they could not automate the entire process and instead just scripted the data processing and display.  Then they confirmed each putative active as a singleton. 
What they are putting together is a "One Size Fits All" process.  I give them credit for doing this, but I think that you cannot find a single NMR-based process for all targets.  In particular, I think they could have used more typical conditions for the reference spectra.  The paper then goes on and discusses their application to targets of interest.  For me, that is irrelevant.  This paper is an excellent companion to the Current Protocol paper, and due to open access, most likely to get far more citations.