Mass spectrometry (MS) is one of the less common tools to
find fragments. In the conceptually simplest approach (native mass spectrometry), you incubate your protein with a putative ligand and ionize the
mixture. Fragment binding is detected by an increased mass for the complex, and the strength of binding
by the ratio of heavier bound complex peak to protein peak. However, the liquid
to gas phase transition is a big step, and often the complex does not survive. Aside
from more specialized applications of MS (such as here, here, and here) there
aren’t many published examples. A recent paper from Federico Sirtori and colleagues
at Nerviano and Università degli Studi di Milano in Eur. J. Pharm. Sci. describes fragment screening by native MS in detail.
The researchers used the reliable model protein Hsp90, which
was also used in a previous MS study and in benchmarking other techniques. One
of the many benefits of Hsp90 is a wealth of well-characterized inhibitors with
a range of affinities, and these were used to calibrate the technique. This
turned out to be critical: beyond sample preparation itself (beware
non-volatile buffer components), all kinds of parameters can be adjusted
including various voltages, temperatures, vacuum strength, and ion source. Get
one of these wrong and your non-covalent complex either fails to ionize or
blows apart.
In addition to using published data on known compounds, the
researchers ran both fluorescence polarization (FP) and surface plasmon
resonance (SPR) assays to independently determine dissociation constants.
Initially the results from MS (a Q-TOF) were quite different, but after
optimization the team was ultimately able to find conditions that gave
qualitatively as well as quantitatively similar results for ligands with
affinities ranging from picomolar to ~100 micromolar.
Thus encouraged, the team embarked on a fragment screening
campaign. The Nerviano fragment library consists of 1914 molecules mostly
following the rule of 3, though halogenated fragments up to 380 Da are allowed
as are compounds with up to 6 hydrogen bond acceptors. The fragments were run
in mixtures of 5, with protein at 2.5 µM and each compound at the low
concentration of 10 µM. Sample injection and data processing were automated,
and the entire screen took 2 days and 2 mg of protein.
Given the low concentration of fragments, the researchers
lowered the bar for potential hits, yielding 282 compounds. These were retested
individually, yielding 146 confirmed hits that gave signals of 5.2-29.7% bound
protein. This is a high hit-rate, particularly given that these binding levels
suggest affinities in the 20-179 µM range. Indeed, only 5 fragments could be competed by
a high-affinity binder, suggesting either that the others bind outside the
active site or are non-specific (false positives). Regarding false negatives, Nerviano
reported the results of an NMR fragment screen against Hsp90 last year, and 12
of 14 hits identified there could also be detected by MS. The other two were
likely below the detection limit of the MS assay.
Unfortunately, the researchers do not discuss thermodynamics.
In theory enthalpic interactions dominate over entropic interactions in the gas
phase, but it is unclear whether any of the observed binders were strongly
entropy-driven.
In the end, it appears that fragment screening by native MS
is workable, but the sensitivity is probably lower than other techniques. Of
course, increasing the ligand concentration would increase the sensitivity to
weaker binders, but at the cost of more non-specific binding – which is already
considerable. Also, Hsp90 is about the friendliest protein one can imagine. I
would be reluctant to try this with a more challenging target that lacks good
tool ligands. But if you want to give it a go, this paper provides a wealth of
information for getting started. And if you have experience with native MS,
please share it in the comments.
3 comments:
I really really like native MS. I have seen it work where SPR outright fails. There are some tricks to the trade of course. It has some significant advantages over the other direct detect biophysical methods: protein QC every experiment and the ability to do stoichiometry for example.
Non-specific adduction occurring during electrospray has been largely documented by Klassen's group (see a nice review from Kitova et al. published in JASMS (2012) "Reliable determination of Protein-Ligand interactions by direct ESI-MS measurements. Are we there yet?"). Artefactual adduction of compound to protein during solvent evaporation, concentration and fission of the charged dropplets typically show Poisson-like distribution. This phenomenum is prone to occur with large excess of compound in incubation mixtures.
Therefore screening for low affinity hits, in the three digit µM range for example (typically fragment molecules), using native MS, absolutely requires high affinity tool molecule to be available to distinguish true hits from false positives. From our experience a screen of 1500 fragment molecules (MW ~ 200) against a nuclear hormone receptor target produced about 3.5% validated hits. Binding stoichiometry and reversibility as well as an estimation of the binding affinity range were also directly available from mass spectra. Screen was done side by side against the apo receptor and against the receptor saturated with the tool compound.
Regarding false negative due to in source dissociation of the complex, the group of John Klassen has described additives and indirect approach that may be beneficial (see above review).
Thanks Teddy and Valerie. I do like the idea of native MS, I just don't recall seeing many examples where it excels in true discovery mode as opposed to proof of concept.
It's true that native MS does give you protein QC, though this can be a limitation if you're working with, for example, glycosylated protein.
Also, you can get stoichiometry with SPR, though it requires more effort. That said, I wonder how often apparent stoichiometry observed with native MS is actually just non-specific binding.